9095-10114-115 PeelASeal Foil Super

9095-10114-115 PeelASeal Foil Super

PeelASeal Foil Super (115mm wide rolls)

Description
A 'stick to all' peelable, foil laminate heat-seal film which is suited for all plate types - Polypropylene (PP), Polystyrene (PS) and Cyclo Olefin Copolymer (COC). The film has a good liquid barrier and high resistance to solvents. It is peelable (from -80℃ freezer) and is non-pierceable. This seal has a white colour to the top aspect.

PRODUCT CODESEAL TYPESIZE
9095-10114-115LRPeelASeal Foil Super500M x 115mm
9095-10114-115SRPeelASeal Foil Super - Sterile500M x 115mm
9095-10114-115TRPeelASeal Foil Super5M x 115mm

Compatibility
Polypropylene (PP), Polystyrene (PS) & Cyclo Olefin Copolymer (COC) plates.

Application
PCR, low temperature, short term compound storage, short term room temperature compound storage (less than 5 days).

Storage
Store in a cool place. Avoid direct exposure to sunlight. It is recommended to use the seals within three years from date of purchase. Three years when stored at 21℃ (70℉), 50% relative humidity, out of direct sunlight, in original packaging.

Properties
Temperature range -80℃ to 110℃.

Sealing
Temperature and Dwell Time: 175℃, 2 seconds.
Recommended sealing equipment: Ultraseal Lite / PRO / XT PRO / ePRO, Wasp, Chameleon, Kube, ALPS3000/5000, A4S, PlateLoc.

SPECIFICATIONS
Visual Description

Metallic with upper surface gloss white. Seal surface metallic burnished foil.

Physical Properties
Flexible, not easily creased. Thicker than IST-104. Temperature Range: -80℃ to +110℃.

TEST PROCEDURES
Mass Loss

Confirming the materials ability to resist high temperatures. Results: Pass.
Details: Mass loss of solution evaluated after 30 cycles of 3 step PCR Programme. Equipment: ABI Thermocycler, Precision Balance.

Pierce
Measuring the force required to push a standardised needle through the material via compression measuring equipment. Results: N/A.
Details 5 tests run using a standardised needle, ensuring that less than 10N is required to pierce the surface & access the wells. Equipment Instron 3343 Tensometer.

Optical
Determining the materials optical clarity by measuring the transmission of emissive dye through the material. Results N/A.
Details Record the light transmission of a sealed microplate using a Flurophore dye stock solution and a microplate reader. Equipment BMG Labtech - FluroStar.

Peel
Measuring the materials permanence of adhesion & its ability to be removed, via extension measuring equipment. Results: Pass.
Details Cohesive Failure, Adhesive Transfer, Material tear & Successful Peel are measured & recorded after a 180℃ peel test. Equipment Instron 3343 Tensometer.

Low Temperature Seal Test
Confirming the materials ability to breath. Results: Pass.
Details: Microplates are sealed at specified low temperatures & subjected to a series of tests to substantiate seal integrity. Equipment: Laboratory Cold storage unit.

Solvent
Evaluating the materials resistance to solvents (DMSO used as an aggressive standard). Results: Pass.
Details: Sealed plate is subjected to a high concentration of DMSO for a time period at low temperatures after which seal damage & volume loss are determined. Equipment Laboratory Cold storage unit, DMSO solution.

Plate Types, Sealing Temp. Time Settings
Polypropylene (PP), Polyethylene (PE), Polystyrene (PS), Cyclo Olefin Copolymer (COC) and non-binding coated plates.
Temperature and Dwell Time: 175℃, 2 seconds.

QC testing is carried out to ensure that products are free from nucleases (DNases & RNases) as well as human genomic DNA. Although every effort is made, including cleanroom manufacture, to maintain this level of cleanliness, best laboratory practice with regards to duplicate testing should be followed.

Kbio back button