9095-10113-078 Gas PermASeal 2

9095-10113-078 Gas PermASeal 2

Gas PermASeal 2 (78mm wide rolls)

Description
A 60gsm Paper with a grid lacquer coating to give a smooth peel, the Seal is Porous, Gas Permeable and a Barrier to Solid Contaminants. It seals to Polypropylene and Polystyrene plates. The seal is pierceable and peelable, and not certified free from nucleases and DNA.

PRODUCT CODESEAL TYPESIZE
9095-10113-078LRGas PermASeal 2610M x 78mm
9095-10113-078SRGas PermASeal 2 - Sterile610M x 78mm
9095-10113-078TRGas PermASeal 25M x 78mm

Compatibility
Polypropylene (PP), Polystyrene (PS).

Application
Short term incubation, agriculture and seed storage, insect storage, cell culture.

Storage
Store in a cool place. Avoid direct exposure to sunlight. It is recommended to use the seals within three years from date of purchase. Three years when stored at 21℃ (70℉), 50% relative humidity, out of direct sunlight, in original packaging.

Properties
Temperature range -20℃ to 80℃.

Sealing
Temperature and Dwell Time: 175℃, 2 seconds.
Recommended sealing equipment: Ultraseal Lite / PRO / XT PRO / ePRO, Wasp, Chameleon, Kube, ALPS3000/5000, A4S, PlateLoc.

SPECIFICATIONS
Visual Description

Upper 60gsm paper. Seal side grid effect lacquer coating.

Physical Properties
Temperature Range: -20℃ to +80℃. Compatibility: Polypropylene (PP), Polystyrene (PS).

TEST PROCEDURES
Mass Loss

Confirming the materials ability to resist high temperatures. Results: N/A.
Details: Mass loss of solution evaluated after 30 cycles of 3 step PCR Programme. Equipment: ABI Thermocycler, Precision Balance.

Pierce
Measuring the force required to push a standardised needle through the material via compression measuring equipment. Results: Pass.
Details 5 tests run using a standardised needle, ensuring that less than 10N is required to pierce the surface & access the wells. Equipment Instron 3343 Tensometer.

Optical
Determining the materials optical clarity by measuring the transmission of emissive dye through the material. Results N/A.
Details Record the light transmission of a sealed microplate using a Flurophore dye stock solution and a microplate reader. Equipment BMG Labtech - FluroStar.

Peel
Measuring the materials permanence of adhesion & its ability to be removed, via extension measuring equipment. Results: Pass.
Details Cohesive Failure, Adhesive Transfer, Material tear & Successful Peel are measured & recorded after a 180℃ peel test. Equipment Instron 3343 Tensometer.

Porosity Bendsten
Confirming the materials ability to breath. Results: Pass.
Details: Measure the defined volume of air forced through the material by specified pressure. Test Method: ISO3781, Units ml/min Target: 25.

Solvent
Evaluating the materials resistance to solvents (DMSO used as an aggressive standard). Results: Pass.
Details: Sealed plate is subjected to a high concentration of DMSO for a time period at low temperatures after which seal damage & volume loss are determined. Equipment Laboratory Cold storage unit, DMSO solution.

Plate Types, Sealing Temp. Time Settings
Polypropylene (PP) Polystyrene (PS) Cyclo Olefin Copolymer (COC).
Temperature and Dwell Time: 175℃, 2 seconds.

QC testing is carried out to ensure that products are free from nucleases (DNases & RNases) as well as human genomic DNA. Although every effort is made, including cleanroom manufacture, to maintain this level of cleanliness, best laboratory practice with regards to duplicate testing should be followed.

Kbio back button