9095-10113-078 (sheets) Gas PermASeal 2
Gas PermASeal 2 (cut sheets)
Description
A 60gsm Paper with a grid lacquer coating to give a smooth peel, the Seal is Porous, Gas Permeable and a Barrier to Solid Contaminants. It seals to Polypropylene and Polystyrene plates. The seal is pierceable and peelable, and not certified free from nucleases and DNA.
A 60gsm Paper with a grid lacquer coating to give a smooth peel, the Seal is Porous, Gas Permeable and a Barrier to Solid Contaminants. It seals to Polypropylene and Polystyrene plates. The seal is pierceable and peelable, and not certified free from nucleases and DNA.
PRODUCT CODE | SEAL TYPE | SHEETS | SIZE |
---|---|---|---|
9095-10113-078LS | Gas PermASeal 2 | 100 | 125mm x 78mm |
9095-10113-078SS | Gas PermASeal 2 - Sterile | 100 | 125M x 78mm |
9095-10113-078TS | Gas PermASeal 2 | 5 | 125M x 78mm |
Compatibility
Polypropylene (PP), Polystyrene (PS).
Application
Short term incubation, agriculture and seed storage, insect storage, cell culture.
Storage
Store in a cool place. Avoid direct exposure to sunlight. It is recommended to use the seals within three years from date of purchase. Three years when stored at 21℃ (70℉), 50% relative humidity, out of direct sunlight, in original packaging.
Properties
Temperature range -20℃ to 80℃.
Sealing
Temperature and Dwell Time: 175℃, 2 seconds.
Recommended sealing equipment: Ultraseal Lite / PRO / XT PRO / ePRO, Wasp, Chameleon, Kube, ALPS3000/5000, A4S, PlateLoc.
SPECIFICATIONS
Visual Description
Upper 60gsm paper. Seal side grid effect lacquer coating.
Physical Properties
Temperature Range: -20℃ to +80℃. Compatibility: Polypropylene (PP), Polystyrene (PS).
TEST PROCEDURES
Mass Loss
Confirming the materials ability to resist high temperatures. Results: N/A.
Details: Mass loss of solution evaluated after 30 cycles of 3 step PCR Programme. Equipment: ABI Thermocycler, Precision Balance.
Pierce
Measuring the force required to push a standardised needle through the material via compression measuring equipment. Results: Pass.
Details 5 tests run using a standardised needle, ensuring that less than 10N is required to pierce the surface & access the wells. Equipment Instron 3343 Tensometer.
Optical
Determining the materials optical clarity by measuring the transmission of emissive dye through the material. Results N/A.
Details Record the light transmission of a sealed microplate using a Flurophore dye stock solution and a microplate reader. Equipment BMG Labtech - FluroStar.
Peel
Measuring the materials permanence of adhesion & its ability to be removed, via extension measuring equipment. Results: Pass.
Details Cohesive Failure, Adhesive Transfer, Material tear & Successful Peel are measured & recorded after a 180℃ peel test. Equipment Instron 3343 Tensometer.
Porosity Bendsten
Confirming the materials ability to breath. Results: Pass.
Details: Measure the defined volume of air forced through the material by specified pressure. Test Method: ISO3781, Units ml/min Target: 25.
Solvent
Evaluating the materials resistance to solvents (DMSO used as an aggressive standard). Results: Pass.
Details: Sealed plate is subjected to a high concentration of DMSO for a time period at low temperatures after which seal damage & volume loss are determined. Equipment Laboratory Cold storage unit, DMSO solution.
Plate Types, Sealing Temp. Time Settings
Polypropylene (PP) Polystyrene (PS) Cyclo Olefin Copolymer (COC).
Temperature and Dwell Time: 175℃, 2 seconds.
QC testing is carried out to ensure that products are free from nucleases (DNases & RNases) as well as human genomic DNA. Although every effort is made, including cleanroom manufacture, to maintain this level of cleanliness, best laboratory practice with regards to duplicate testing should be followed.