9095-10110-078 (sheets) Gas PermASeal

9095-10110-078 (sheets) Gas PermASeal

Gas PermASeal (cut sheets)

Description
An opaque, non-woven porous and gas permeable film which acts as a barrier to solid contaminants. It seals to Polypropylene and Polystyrene plates. The seal is pierceable and peelable, and not certified free from nucleases and DNA.

PRODUCT CODESEAL TYPESHEETSSIZE
9095-10110-078LSGas PermASeal100125mm x 78mm
9095-10110-078SSGas PermASeal - Sterile100125M x 78mm
9095-10110-078TSGas PermASeal5125M x 78mm

Compatibility
Polypropylene (PP), Polystyrene (PS).

Application
Short term incubation, agriculture and seed storage, insect storage, cell culture.

Storage
Store in a cool place. Avoid direct exposure to sunlight. It is recommended to use the seals within three years from date of purchase. Three years when stored at 21℃ (70℉), 50% relative humidity, out of direct sunlight, in original packaging.

Properties
Temperature range -20℃ to 80℃

Sealing
Temperature and Dwell Time: 170℃, 2 seconds.
Recommended sealing equipment: Ultraseal Lite / PRO / XT PRO / ePRO, Wasp, Chameleon, Kube, ALPS3000/5000, A4S, PlateLoc.

SPECIFICATIONS
Visual Description

White non-woven. Seal side has a shiny lacquer coating.

Physical Properties
Temperature Range: -20℃ to +80℃. Compatibility: Polypropylene (PP), Polystyrene (PS).

TEST PROCEDURES
Mass Loss

Confirming the materials ability to resist high temperatures. Results: N/A.
Details: Mass loss of solution evaluated after 30 cycles of 3 step PCR Programme. Equipment: ABI Thermocycler, Precision Balance.

Pierce
Measuring the force required to push a standardised needle through the material via compression measuring equipment. Results: Pass.
Details 5 tests run using a standardised needle, ensuring that less than 10N is required to pierce the surface & access the wells. Equipment Instron 3343 Tensometer.

Optical
Determining the materials optical clarity by measuring the transmission of emissive dye through the material. Results N/A.
Details Record the light transmission of a sealed microplate using a Flurophore dye stock solution and a microplate reader. Equipment BMG Labtech - FluroStar.

Peel
Measuring the materials permanence of adhesion & its ability to be removed, via extension measuring equipment. Results: Pass.
Details Cohesive Failure, Adhesive Transfer, Material tear & Successful Peel are measured & recorded after a 180℃ peel test. Equipment Instron 3343 Tensometer.

Water Vapour Transmission
Confirming the materials ability to breath. Results: Pass.
Details: Measure the weight loss of water during a set time at a set temperature and humidity Test Method: T30/001, Ref ASTM E-96-66, Target: 1800 g/m2/24h.

Solvent
Evaluating the materials resistance to solvents (DMSO used as an aggressive standard). Results: Pass.
Details: Sealed plate is subjected to a high concentration of DMSO for a time period at low temperatures after which seal damage & volume loss are determined. Equipment Laboratory Cold storage unit, DMSO solution.

Plate Types, Sealing Temp. Time Settings
Polypropylene(PP), Polystyrene (PS).
Temperature and Dwell Time: 160℃, 2 seconds.

QC testing is carried out to ensure that products are free from nucleases (DNases & RNases) as well as human genomic DNA. Although every effort is made, including cleanroom manufacture, to maintain this level of cleanliness, best laboratory practice with regards to duplicate testing should be followed.

Kbio back button