9095-10107-115 PierceASeal Foil PS

9095-10107-115 PierceASeal Foil PS

PierceASeal Foil PS (115mm wide rolls)

Description
A pierceable A high grade foil with good solvent resistance including DMSO, easy sealing surface identification with multiple sealing and resealing properties. The seal is peelable Polystyrene only and pierceable. foil seal with easy sealing surface identification and high solvent resistance including DMSO. The film is non-peelable, pierceable and re-sealing is permissible.

PRODUCT CODESEAL TYPESIZE
9095-10107-115LRPierceASeal Foil PS500M x 115mm
9095-10107-115SRPierceASeal Foil PS - Sterile500M x 115mm
9095-10107-115TRPierceASeal Foil PS5M x 115mm

Compatibility
Polypropylene (PP), Polystyrene (PS).

Application
PCR low temperature compound storage, short term room temperature compound storage.

Storage
Store in a cool place. Avoid direct exposure to sunlight. It is recommended to use the seals within three years from date of purchase. Three years when stored at 21℃ (70℉), 50% relative humidity, out of direct sunlight, in original packaging.

Properties
Temperature range -20℃ to 110℃.

Sealing
Temperature and Dwell Time: 175℃, 2 seconds.
Recommended sealing equipment: Ultraseal Lite / PRO / XT PRO / ePRO, Wasp, Chameleon, Kube, ALPS3000/5000, A4S, PlateLoc.

SPECIFICATIONS
Visual Description

Metallic reflective foil, with both sides appearing very similar. Printed line denotes upper surface.

Physical Properties
Very flexible foil, not easily creased. Temperature Range: -20℃ to 110℃.

TEST PROCEDURES
Mass Loss

Confirming the materials ability to resist high temperatures. Results: Pass.
Details: Mass loss of solution evaluated after 30 cycles of 3 step PCR Programme. Equipment: ABI Thermocycler, Precision Balance.

Pierce
Measuring the force required to push a standardised needle through the material via compression measuring equipment. Results: Pass.
Details 5 tests run using a standardised needle, ensuring that less than 10N is required to pierce the surface & access the wells. Test Value = 7.22N. Equipment Instron 3343 Tensometer.

Burst Testing
Determining the materials adhesion to the plate. Results Pass.
Details Microplates are sealed and tested under pressure. Tests passed once achieved 2 bar of pressure or greater. Equipment Miniburst 5.

Peel
Measuring the materials permanence of adhesion & its ability to be removed, via extension measuring equipment. Results: Pass.
Details Cohesive Failure, Adhesive Transfer, Material tear & Successful Peel are measured & recorded after a 180℃ peel test. Equipment Instron 3343 Tensometer.

Low Temperature Seal Test
Confirming the materials ability to resist low temperatures. Results: Pass.
Details: Microplates are sealed at specified low temperatures & subjected to a series of tests to substantiate seal integrity. Equipment: Laboratory Cold storage unit.

Solvent
Evaluating the materials resistance to solvents (DMSO used as an aggressive standard). Results: N/A.
Details Sealed plate is subjected to a high concentration of DMSO for a time period at low temperatures after which seal damage & volume loss are determined. Equipment Laboratory Cold storage unit, DMSO solution.

Plate Types, Sealing Temp. Time Settings
Polypropylene(PP), Polystyrene (PS).
Temperature and Dwell Time: 175℃, 2 seconds.

QC testing is carried out to ensure that products are free from nucleases (DNases & RNases) as well as human genomic DNA. Although every effort is made, including cleanroom manufacture, to maintain this level of cleanliness, best laboratory practice with regards to duplicate testing should be followed.

Kbio back button