9095-10105-115 PeelASeal DMSO Foil
PeelASeal DMSO Foil (115mm wide rolls)
Description
A DMSO resistant foil laminate suited for Polypropylene plates, with a good liquid barrier and high solvent-resistance (at high temperatures). The seal is peelable and non-pierceable
A DMSO resistant foil laminate suited for Polypropylene plates, with a good liquid barrier and high solvent-resistance (at high temperatures). The seal is peelable and non-pierceable
PRODUCT CODE | SEAL TYPE | SIZE |
---|---|---|
9095-10105-115LR | PeelASeal DMSO Foil | 500M x 115mm |
9095-10105-115SR | PeelASeal DMSO Foil - Sterile | 500M x 115mm |
9095-10105-115TR | PeelASeal DMSO Foil | 5M x 115mm |
Compatibility
Polypropylene (PP) Plates.Application
Low temperature and ambient temperature storage with DMSO and other solvents.
Storage
Store in a cool place. Avoid direct exposure to sunlight. It is recommended to use the seals within three years from date of purchase. Three years when stored at 21℃ (70℉), 50% relative humidity, out of direct sunlight, in original packaging.
Properties
Temperature range -20℃ to 120℃.
Sealing
Temperature and Dwell Time: 175℃, 2 seconds.
Recommended sealing equipment: Ultraseal Lite / PRO / XT PRO / ePRO, Wasp, Chameleon, Kube, ALPS3000/5000, A4S, PlateLoc.
SPECIFICATIONS
Visual Description
Upper glossy metallic surface. Sealing surface less reflective, more highly burnished and smoother.
Physical Properties
Flexible, not easily creased. Temperature Range: -80℃ to +80℃.
TEST PROCEDURES
Mass Loss
Confirming the materials ability to resist high temperatures. Results: Pass.
Details: Mass loss of solution evaluated after 30 cycles of 3 step PCR Programme. Equipment: ABI Thermocycler, Precision Balance.
Pierce
Measuring the force required to push a standardised needle through the material via compression measuring equipment. Results: N/A.
Details 5 tests run using a standardised needle, ensuring that less than 10N is required to pierce the surface & access the wells. Equipment Instron 3343 Tensometer.
Optical
Determining the materials optical clarity by measuring the transmission of emissive dye through the material. Results: N/A.
Details Record the light transmission of a sealed microplate using a Flurophore dye stock solution and a microplate reader. Equipment BMG Labtech - FluroStar.
Peel
Measuring the materials permanence of adhesion & its ability to be removed, via extension measuring equipment. Results: Pass.
Details Cohesive Failure, Adhesive Transfer, Material tear & Successful Peel are measured & recorded after a 180℃ peel test. Equipment Instron 3343 Tensometer.
Low Temperature Seal Test
Confirming the materials ability to resist low temperatures. Results: Pass.
Details: Microplates are sealed at specified low temperatures & subjected to a series of tests to substantiate seal integrity. Equipment: Laboratory Cold storage unit.
Solvent
Evaluating the materials resistance to solvents (DMSO used as an aggressive standard) Results: N/A Details Sealed plate is subjected to a high concentration of DMSO for a time period at low temperatures after which seal damage & volume loss are determined. Equipment Laboratory Cold storage unit, DMSO solution.
Plate Types, Sealing Temp. Time Settings
Polypropylene (PP), certain Cyclo Olefin Copolymer(COC) plates, welds to Polyethylene (PE).
Details: Sealed plate is subjected to a high concentration of DMSO for a time period at low temperatures after which seal damage & volume loss are determined. Equipment Laboratory Cold storage unit, DMSO solution.
QC testing is carried out to ensure that products are free from nucleases (DNases & RNases) as well as human genomic DNA. Although every effort is made, including cleanroom manufacture, to maintain this level of cleanliness, best laboratory practice with regards to duplicate testing should be followed.