9095-10124-080 (adhesive) QuickSeal Gas Perm

9095-10124-080 (adhesive) QuickSeal Gas Perm

QuickSeal Gas Perm (adhesive seals)

Description
A transparent, perforated gas permeable film. The seal is perforated and permeable to gases. It is peelable and pierceable.

PRODUCT CODESEAL TYPEQUANTITYSIZE
9095-10124-080LRQuickSeal Gas PermRoll100M x 80mm
9095-10124-080SRQuickSeal Gas Perm - SterileRoll100M x 80mm
9095-10124-080LSQuickSeal Gas Perm100 sheets130mm x 80mm
9095-10124-080SSQuickSeal Gas Perm - Sterile2 x (25) sheets130mm x 80mm
9095-10124-080TSQuickSeal Gas Perm5 sheets130mm x 80mm

Compatibility
Polypropylene (PP), Polystyrene (PS) and Cyclo Olefin Copolymer (COC).

Application
Bacterial culture, Eukaryotic cell culture.

Storage
Store in a cool place. Avoid direct exposure to sunlight. It is recommended to use the seals within three years from date of purchase. Three years when stored at 21℃ (70℉), 50% relative humidity, out of direct sunlight, in original packaging.

Properties
Temperature range -20℃ to 80℃.

Sealing
Recommended sealing equipment: Ultraseal COLD-PRO.

SPECIFICATIONS
Visual Description

Transparent, Perforated EVA medical Tape, Plastic, weave textured, with a cream coloured Liner.

Physical Properties
Single coated tape, consisting of a transparent, perforated, hypoallergenic coated, pressure sensitive acrylate adhesive. Temperature range: -20℃ to +80℃.

TEST PROCEDURES
Mass Loss

Confirming the materials ability to resist high temperatures. Results: Pass.
Details: Mass loss of solutions evaluated after 30 Cycles of 3 step PCR Programme.
Equipment: ABI Thermocycler, Precision Balance.

Pierce
Measuring the force required to push a standardised needle through the material via compression measuring equipment. Results: Pass.
Details 5 tests run using a standardised needle, ensuring that less than 10N is required to pierce the surface & access the wells. Equipment Instron 3343 Tensometer.

Optical
Determining the materials optical clarity by measuring the transmission of emissive dye through the material. Results N/A.
Details Record the light transmission of a sealed microplate using a Flurophore dye stock solution and a microplate reader. Equipment BMG Labtech - FluroStar.

Peel
Measuring the materials permanence of adhesion & its ability to be removed, via extension measuring equipment. Results: Pass.
Details Cohesive Failure, Adhesive Transfer, Material tear & Successful Peel are measured & recorded after a 180℃ peel test. Equipment Instron 3343 Tensometer.

Porosity Bendsten
Confirming the materials ability to resist low temperatures. Results: Pass.
Details: MVTR, gms/m2/day. Air Porosity, Gurley 15 sec/100cc/Sq. in.

Low Temperature Seal Test
Confirming the materials ability to breath. Results: Pass.
Details: Microplates are sealed at specified low temperatures & subjected to a series of tests to substantiate seal integrity. Equipment: Laboratory Cold storage unit.

Solvent
Evaluating the materials resistance to solvents (DMSO used as an aggressive standard) Results: N/A Details Sealed plate is subjected to a high concentration of DMSO for a time period at low temperatures after which seal damage & volume loss are determined. Equipment Laboratory Cold storage unit, DMSO solution.

Plate Types
Polypropylene (PP), Polyethylene (PE), Polystyrene (PS) Cyclo Olefin Copolymer (COC).

QC testing is carried out to ensure that products are free from nucleases (DNases & RNases) as well as human genomic DNA. Although every effort is made, including cleanroom manufacture, to maintain this level of cleanliness, best laboratory practice with regards to duplicate testing should be followed.